OPTISYS TUR 1060

Optical turbidity measuring system for potable water applications

  • 4-wire, 4…20 mA, 2 alarm relays or Modbus via RS485; built-in data logger
  • For online bypass installation
  • 0…100/1000 NTU/FNU; max. +50°C / +122°F; 7 bar / 101 psi
  • 90° scattered infrared light (ISO 7027) or white light (US EPA 180.1)

OPTISYS TUR 1060 – Standard version

Turbidity measuring system OPTISYS TUR 1060 – Standard version

The OPTISYS TUR 1060 is a turbidity measuring system with cost-efficient cuvette calibration and automated ultrasonic cleaning system. Due to its optimised cuvette measuring system, it has a very fast measuring response time. The sensor uses the 90° scattered light method, thus the measurement optics are not directly exposed to the sample and require less maintenance.

The low-range system is the most suitable system when it comes to fulfilling all the necessary regulations and requirements in water and wastewater treatment. The built-in data logger allows storage of readings and calibration data. The measuring data can be downloaded via the USB interface. In addition, calibration data can be uploaded via USB to change the measuring range.

Product highlights

  • Rapid response time due to small measurement volume
  • Improved usability due to easy-to-operate keypad
  • Data logging with user-selectable interval (1…60 min) for storage of 1 year’s readings and calibration data
  • USB interface for easy data exchange
  • Simple cuvette calibration with reusable liquid calibration standards
  • Integrated shut-off valve for maintenance work
  • Optimal back pressure setting to avoid gas bubbles through integrated outlet valve
  • Automatic ultrasonic cleaning to prevent deposits

Typical applications

Water industry

  • Monitoring of potable water quality in pumping stations
  • Filter monitoring
  • Process control of water treatment
  • Monitoring of limit values
  • Demineralisation
  • Monitoring of water quality prior to network supply

Applications