OPTIFLEX 3200
Guided radar (TDR) level transmitter for liquids with hygienic requirements
- Continuous measurement of level and interface in small and narrow vessels
- Conform to FDA, EC1935/2004, EHEDG
- Measuring range: 0.6…4 m / 2…13 ft
- -50…+150°C / -58…+302°F; -1…40 barg / -14.5…580 psig
OPTIFLEX 3200 C–Compact version with polished Ø8 / 0.32 mm single rod probe
The OPTIFLEX 3200 is a 2-wire guided radar level transmitter based on the Time Domain Reflectometry (TDR) technology. It is designed for the measurement of level and interface in the food, beverage and pharmaceutical industry. The TDR is insensitive to steam, foam or condensation and suitable for a large variety of liquids with hygienic requirements. The wetted parts have a gap-free hygienic design that ensures simple and reliable cleanability (CIP/SIP). Dedicated industry-specific probes and process connections as well as various converter and electronic versions are available to suit every installation and facilitate device access.
Product highlights
- Hygienic probes and process connections for the food and beverage as well as pharmaceutical industry
- Suitable for SIP (Sterilisation In Place) and CIP (Cleaning In Place)
- Insensitive to steam, foam and condensation
- Measurement of interfaces starting at 50 mm / 1.97"
- Various converter and electronic versions to facilitate access to the device:
- Available as compact version (C) or with remote converter (F) up to 100 m / 328 ft
- Horizontal or vertical housing to suit every installation
- Converter is rotatable and removable under process conditions and compatible with all previous Guided Radar (TDR) level transmitters made by KROHNE
- Cast stainless steel or aluminium housing
- ±2 mm /±0.08" accuracy
- Real-time clock for event logging
- 3-year warranty
Typical applications
General
- Level and interface measurement of liquids in hygienic applications with finished and raw products
- Applications with storage tanks, vessels, mixers and containers
Food and beverage industry
- Fruit extract, compote
- Raw milk, cheese
- Beer, wine
- Yeast
- Fruit juice, tomato sauce, soup
- Vegetable oil, palm oil
- Alcohols, purified water
Pharmaceutical industry
- Solvents, alcohols, aqueous solutions
- Slightly corrosive acids, alkalis
- Chlorobenzene, chloroform
- Vaccines, plasma
Applications
General information
Measurement principle | Guided radar |
Level measurement type | Continuous measurement |
Measured media | Liquid/liquid interface, Liquids |
Contact to measured medium | Yes |
Interface measurement | Yes |
Probe diameter | |
Probe length | 0.6...4 m / 1.97...13.12 ft |
Measuring range (liquids) | 0.6…4 m / 1.97…13.12 ft |
Process pressure | -1…40 barg / -14,5…580 psig |
Process temperature | -50…+150°C / -58…+302°F |
Wetted parts | Austenitic stainless steel (1.4404/316L) |
Gaskets | Ceramic, Ethylene propylene diene monomer (EPDM), Fluoroelastomer rubber (FKM/FPM), Kalrez® 6221, PEEK, Polytetrafluoroethylene (PTFE) |
Accuracy (liquids) | ±2 mm / ±0,08" |
Approvals/Standards
Certificates/Approvals Ex | ATEX, cQPSus, IECEx, INMETRO, NEPSI |
Certificates/Approvals Safety | SIL 2 |
Certificates/Approvals Hygienic | EC 1935/2004, EHEDG, FDA |
Ingress protection | IP66/68, NEMA4X/6P |
Communication
Digital outputs | HART® |
Discrete outputs | Relay |
Analogue outputs | 4…20 mA |
Process connections
Hygienic connections | Dairy screw (DIN 11851): DN25, DN40, DN50 SMS (1145): 38 mm, 51 mm Clamp (DIN 32676): DN25, DN40, DN50 Clamp (ISO 2852): 25, 38, 51 Tri-Clamp®: 1", 11⁄2", 2" |
Download Center
OPTIFLEX 3200
Download Center
- Technical datasheets and handbooks
- Flyers and brochures
- Certificates and approvals
- Software and driver